ORTHO-CLINICAL DIAGNOSTICS recalls VITROS¿ Immunodiagnostic Products CA 19-9 Calibrators and Reagent Packs, Lot 1320 VITROS¿ CA 19-9 Reagent Packs are u
Recalled August 20, 2014 · FDA recall Z-2183-2014 · ORTHO-CLINICAL DIAGNOSTICS
Hazard
Customers may receive positively biased results when using VITROS¿ Immunodiagnostic Products CA 19-9 Total Reagent Packs, Lot 1320 and VITROS¿ CA 19-9 Calibrators, Lot 1320 on the VITROS ECi/ECiQ Immunodiagnostic Systems, the VITROS 3600 Immunodiagnostic System and the VITROS 5600 Integrated System. There is an unexpected observed change in CA 19-9 Calibrators, Lot 1320 calibrator values. · FDA Class II
Units
475
Sold at
Worldwide Distribution - USA (nationwide) and Internationally to France and Japan
Description
Customers may receive positively biased results when using VITROS¿ Immunodiagnostic Products CA 19-9 Total Reagent Packs, Lot 1320 and VITROS¿ CA 19-9 Calibrators, Lot 1320 on the VITROS ECi/ECiQ Immunodiagnostic Systems, the VITROS 3600 Immunodiagnostic System and the VITROS 5600 Integrated System. There is an unexpected observed change in CA 19-9 Calibrators, Lot 1320 calibrator values.
Product
VITROS¿ Immunodiagnostic Products CA 19-9 Calibrators and Reagent Packs, Lot 1320 VITROS¿ CA 19-9 Reagent Packs are used on VITROS¿ ECi/ECiQ Immunodiagnostic Systems, VITROS¿ 3600 Immunodiagnostic Systems and VITROS¿ 5600 Integrated Systems for the quantitative measurement of 1116-NS-19-9 defined antigen in human serum and plasma.
Manufacturer(s)
ORTHO-CLINICAL DIAGNOSTICS
Made in
United Kingdom
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →