Home / Recalls / Medical Devices
Ortho-Clinical Diagnostics recalls VITROS Immunodiagnostic Products CK-MB Calibrators, REF 128 6293, IVD, Ortho-Clinical Diagnostics High Wycombe HP12 4DP
Recalled May 29, 2013 · FDA recall Z-1337-2013 · Ortho-Clinical Diagnostics
Hazard
Internal testing demonstrated that opened-reconstituted VITROS CK-MB Calibrators stored at -20 degrees Celsius (-4 degrees Farenheit) generated invalid calibrations that were detected by positively biased quality control results. · FDA Class III
Units
Domestic: 3019 units, Foreign: 905 units
Sold at
Worldwide Distribution - Nationwide including Puerto Rico, and the countries of Australia, Bermuda, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Singapore, Spain and Venezuela.
Description
Internal testing demonstrated that opened-reconstituted VITROS CK-MB Calibrators stored at -20 degrees Celsius (-4 degrees Farenheit) generated invalid calibrations that were detected by positively biased quality control results.
Product
VITROS Immunodiagnostic Products CK-MB Calibrators, REF 128 6293, IVD, Ortho-Clinical Diagnostics High Wycombe HP12 4DP, UK For use in the calibration of the VITROS ECi/ECiQ Immunodiagnostic Systems, the VITROS 3600 Immunodiagnostic System and the VITROS 5600 Integrated System for the quantitative measurement of CK-MB in human serum and plasma (EDTA or heparin).
Manufacturer(s)
Ortho-Clinical Diagnostics
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →