Ortho-Clinical Diagnostics recalls VITROS Performance Verifier I
Recalled April 3, 2024 · FDA recall Z-1372-2024 · Ortho-Clinical Diagnostics
Hazard
The issue was mismatched diluent lots assembled into VITRO Chemistry Products Performance Verifier I and II. This issue has the potential to cause erroneous Quality Control (QC) results that could lead to testing delays and erroneous patient results. This issue caused positive or negative bias in the quantification of sodium (Na+) and total carbon dioxide (ECO2). The incorrect diluent into the Performance Verifier II exhibited a potential negative bias of up to 6 mmol/L for ECO2 and Na+ testing. · FDA Class II
Units
10,911 units
Sold at
Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Norway, Portugal, Russia, Singapore, Spain, Sweden, The Netherlands, United Kingdom.
Description
The issue was mismatched diluent lots assembled into VITRO Chemistry Products Performance Verifier I and II. This issue has the potential to cause erroneous Quality Control (QC) results that could lead to testing delays and erroneous patient results. This issue caused positive or negative bias in the quantification of sodium (Na+) and total carbon dioxide (ECO2). The incorrect diluent into the Performance Verifier II exhibited a potential negative bias of up to 6 mmol/L for ECO2 and Na+ testing.
Product
VITROS Performance Verifier I
Manufacturer(s)
Ortho-Clinical Diagnostics
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →