Orthofix, Inc recalls Firebird NXG Spinal Fixation System Set Screw Driver used to insert a set screw into a spinal fixation construct. Rx On
Recalled June 8, 2016 · FDA recall Z-1881-2016 · Orthofix
Hazard
There is a possibility that the subject set screw driver could disengage from the torque limiting handle when the counter torque wrench is allowed to contact the counter torque handle. · FDA Class II
Units
36 units
Sold at
KS, NV, TX, CA, ID, FL, MA, MI, NJ, AZ, NH, SC, UT, WA
Description
There is a possibility that the subject set screw driver could disengage from the torque limiting handle when the counter torque wrench is allowed to contact the counter torque handle.
Product
Firebird NXG Spinal Fixation System Set Screw Driver used to insert a set screw into a spinal fixation construct. Rx Only Provided non Sterile.
Manufacturer(s)
Orthofix, Inc
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →