Home / Recalls / Medical Devices
Orthofix, Inc recalls Lapidus Plate, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Size/Step/Left or Right (Part Number): a) Small/0MM S
Recalled June 13, 2018 · FDA recall Z-2129-2018 · Orthofix
Hazard
Firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery. · FDA Class II
Units
156
Sold at
U.S Consignees: NJ, FL, PA, SC, LA, VA, MN, TX, GA, DC, NV, IN, CA, MI, IL, MD, NC, OH, OK,
Description
Firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.
Product
Lapidus Plate, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Size/Step/Left or Right (Part Number): a) Small/0MM Step/Left (002-C-01000L); b) Small/0MM Step/Right (002-C-01000R); c) Large/0MM Step/Left (002-C-02000L); d) Large/0MM Step/Right (002-C-02000R); e) Small/2MM Step/Left (002-C-01002L); f) Small/2MM Step/Right (002-C-01002R); g) Large/2MM Step/Left (002-C-02002L); h) Large/2MM Step/Right (002-C-02002R)
Manufacturer(s)
Orthofix, Inc
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →