Orthohelix Surgical Designs Inc recalls Intraosseous Fixation System (IFS) 1.7 mm, 2.0 mm & 2.5 mm Cannulated Drill Bit. Intended to stabilize and aid in fix

Recalled April 3, 2013 · FDA recall Z-0975-2013 · Orthohelix Surgical Designs

Hazard
The firm was notified by a customer that there was a burr on the inside end of the drill bit resulting in the drill bit not fitting over the associated guide wire. · FDA Class II
Units
Lot # 2399041, 97 pcs; Lot # 2399051, 143 pcs; Lot # 2399051R, 4 pcs & Lot # 2399061, 119 pcs
Sold at
Nationwide Distribution including the states of CA, CO, IL, MI, MD, NC, NM, OH, OK, PA, SD, TX, UT, VA, WA and WI.

Description

The firm was notified by a customer that there was a burr on the inside end of the drill bit resulting in the drill bit not fitting over the associated guide wire.

Product
Intraosseous Fixation System (IFS) 1.7 mm, 2.0 mm & 2.5 mm Cannulated Drill Bit. Intended to stabilize and aid in fixation of fractures.
Manufacturer(s)
Orthohelix Surgical Designs Inc