RecallWatchUSA

Home / Recalls / Medical Devices

Orthoscan, Inc. recalls Ziehm Vision system, x-ray, tomography, computed.

Recalled December 11, 2024 · FDA recall Z-0575-2025 · Orthoscan

Hazard
Ziehm Imaging, Inc., distributor, have been informed by manufacturer, Ziehm Imaging GmbH that installed devices may not fully meet the requirements of IEC 60601-2-54 and or 21 CFR 1020.32(c) with respect to wired hand switch may fail when being exposed to mechanical stress such as shocks (e.g. dropping) or impacts can irreversibly damage the switching element in the hand switch. A defective switching element may then potentially initiate unintended radiation of the connected mobile X-ray unit. · FDA Class II
Units
3
Sold at
U.S. (including Puerto Rico).

Description

Ziehm Imaging, Inc., distributor, have been informed by manufacturer, Ziehm Imaging GmbH that installed devices may not fully meet the requirements of IEC 60601-2-54 and or 21 CFR 1020.32(c) with respect to wired hand switch may fail when being exposed to mechanical stress such as shocks (e.g. dropping) or impacts can irreversibly damage the switching element in the hand switch. A defective switching element may then potentially initiate unintended radiation of the connected mobile X-ray unit.

Product
Ziehm Vision system, x-ray, tomography, computed.
Manufacturer(s)
Orthoscan, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Orthoscan, Inc. recalls Ziehm Vision system, x-ray, tomography, computed. | RecallWatchUSA