Ossur Americas recalls Ossur, J Tong, Sterile, Part/ Description: JT-115/ J TONG MED/LGE-No App; JT-200/ J TONG ASSEMBLY SML/MED; JT-210

Recalled March 20, 2019 · FDA recall Z-1002-2019 · Ossur Americas

Hazard
Halo system products labeled MR conditional will undergo confirmatory safety testing to confirm MR conditional parameters. In addition, the second line description on the titanium pin label mistakenly lists ceramic tip. The second line description should state titanium pin. · FDA Class II
Units
138
Sold at
U.S.: HI, AL, CA, WA, AZ, TX, NM, NE, MI, IL, MN, MS, IN, OH, KY, TN, AK, AL, UT, SC, NC, GA, PA, FL, NY, DE, RI, VA, MA; Foreign: Australia, Germany, New Zealand, Belgium

Description

Halo system products labeled MR conditional will undergo confirmatory safety testing to confirm MR conditional parameters. In addition, the second line description on the titanium pin label mistakenly lists ceramic tip. The second line description should state titanium pin.

Product
Ossur, J Tong, Sterile, Part/ Description: JT-115/ J TONG MED/LGE-No App; JT-200/ J TONG ASSEMBLY SML/MED; JT-210/ J TONG ONLY SML/MED; JT-400/ J TONG COMPONENT TRAY
Manufacturer(s)
Ossur Americas