Paltop Advanced Dental Solutions Ltd recalls Paltop Dynamic Dental Implant Dia 3.25mm L 13.0 mm, Catalog Number 21-70020

Recalled January 29, 2020 · FDA recall Z-0836-2020 · Paltop Advanced Dental Solutions

Hazard
Some of the implants were packaged with a Standard Platform Cover Screw instead of a Narrow Platform Cover Screw. The Standard Platform Cover Screw will not fit in the Narrow Platform implant. · FDA Class II
Units
67
Sold at
The products were distributed to the following US states: CA, DC, FL, GA, MA, MD, NJ, NY, SC, TX, and VA.

Description

Some of the implants were packaged with a Standard Platform Cover Screw instead of a Narrow Platform Cover Screw. The Standard Platform Cover Screw will not fit in the Narrow Platform implant.

Product
Paltop Dynamic Dental Implant Dia 3.25mm L 13.0 mm, Catalog Number 21-70020
Manufacturer(s)
Paltop Advanced Dental Solutions Ltd
Made in
Israel