RecallWatchUSA

Home / Recalls / Medical Devices

Pega Medical Inc. recalls Slipped Capital Femoral Epiphysis (SCFE) Drivers of the Free-Gliding SCFE Screw System with catalogue number SCF-MLD265

Recalled July 8, 2015 · FDA recall Z-1848-2015 · Pega Medical

Hazard
SCFE Drivers of the Free-Gliding SCFE Screw System are recalled because the handle of the instruments may detach from the shaft if high torques are applied in counter-clockwise rotation. · FDA Class II
Units
4 units
Sold at
US Distribution to the states of : FL and TN.

Description

SCFE Drivers of the Free-Gliding SCFE Screw System are recalled because the handle of the instruments may detach from the shaft if high torques are applied in counter-clockwise rotation.

Product
Slipped Capital Femoral Epiphysis (SCFE) Drivers of the Free-Gliding SCFE Screw System with catalogue number SCF-MLD265 and SCF-MLD273.
Manufacturer(s)
Pega Medical Inc.
Made in
Canada

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov