Pentax Medical Company recalls Laryngo Stroboscope Product Usage Intended to provide optical visualization of, therapeutic access to, the Upper Airway

Recalled July 6, 2016 · FDA recall Z-2073-2016 · Pentax Medical

Hazard
Pentax has become aware that operational/cleaning accessories and therapeutic devices (e.g., clips, stents, balloons, etc.) can become lodged in an endoscope's instrument channel. · FDA Class II
Units
206 units
Sold at
Worldwide Distribution - US (Nation Wide) including Puerto Rico and Internationally to Denmark, Guam

Description

Pentax has become aware that operational/cleaning accessories and therapeutic devices (e.g., clips, stents, balloons, etc.) can become lodged in an endoscope's instrument channel.

Product
Laryngo Stroboscope Product Usage Intended to provide optical visualization of, therapeutic access to, the Upper Airway and Vocal Folds. This anatomy includes, but is not restricted to, the organs, tissues and subsystems: Larynx, Pharynx and Vocal Cords. The instrument is introduced via the mouth or the nose, as decided by the physician, when indications consistent with the need for procedure are observed in adult and pediatric patient populations.