RecallWatchUSA

Home / Recalls / Medical Devices

Pentax Medical Company recalls PENTAX Endoscopic Accessories. OF-B194. Made in Japan. PENTAX Upper/Lower G.I. Endoscopes are provided with Air/Wat

Recalled November 6, 2013 · FDA recall Z-0084-2014 · Pentax Medical

Hazard
PENTAX Medical has become aware of a product issue associated with the OF-B194 Gas/Water Feeding Valve. A manufacturing defect may cause improper stoppage of CO2 gas during an endoscopic procedure with the potential for serious hazard to and/or user. · FDA Class I
Units
76 units
Sold at
Nationwide Distribution including Arizona, California, Florida, Idaho, Indiana, Kansas, Maryland, Missouri, Missippi, North Carolina, New York, Pennsylvania, Virginia, Wisconsin and Wyoming.

Description

PENTAX Medical has become aware of a product issue associated with the OF-B194 Gas/Water Feeding Valve. A manufacturing defect may cause improper stoppage of CO2 gas during an endoscopic procedure with the potential for serious hazard to and/or user.

Product
PENTAX Endoscopic Accessories. OF-B194. Made in Japan. PENTAX Upper/Lower G.I. Endoscopes are provided with Air/Water Valve. As an alternative, the optionally available Gas/Water Valve may be used in place of the standard Air/Water Valve to deliver a non-explosive gas for insufflation.
Manufacturer(s)
Pentax Medical Company

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Pentax Medical Company recalls PENTAX Endoscopic Accessories. OF-B194. Made in Japan. PENTAX Upper/Lower G.I. Endoscopes are provided with Air/Wat | RecallWatchUSA