PENTAX of America Inc recalls C2 CryoBalloon Controller, REF: FG-1017, with Controller Software v1.18.258 The C2 CryoBalloon Ablation System is int
Recalled March 13, 2019 · FDA recall Z-0970-2019 · PENTAX of America
Hazard
The Controller does not detect overpressure in the balloon during the application of non-dosing puffs of Nitrous Oxide, which can contribute to balloon over pressurization, if the intended vent lumen of the catheter is significantly occluded to prevent relieving balloon pressure due to a kinked catheter condition.If a patient is exposed to higher than physiologic pressures, adverse events such as perforation or mucosal laceration may occur. · FDA Class II
Units
45
Sold at
Worldwide Distribution: US (nationwide) to states of:: CA, DC, FL, LA, MA, MD, MN, NC, NJ, NY, OH, PA, SC, TN, TX, VA, WA;; and countries: Canada, Germany, and Netherlands.
Description
The Controller does not detect overpressure in the balloon during the application of non-dosing puffs of Nitrous Oxide, which can contribute to balloon over pressurization, if the intended vent lumen of the catheter is significantly occluded to prevent relieving balloon pressure due to a kinked catheter condition.If a patient is exposed to higher than physiologic pressures, adverse events such as perforation or mucosal laceration may occur.
Product
C2 CryoBalloon Controller, REF: FG-1017, with Controller Software v1.18.258 The C2 CryoBalloon Ablation System is intended to be used as a cryosurgical tool in the field of general surgery, specifically for endoscopic application and the ablation of dysplastic Barrett s Esophagus.
Manufacturer(s)
PENTAX of America Inc
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →