Pentax of America Inc recalls Pentax Medical Video Processor- Intended to be used with PENTAX Medical endoscopes, monitors and other peripheral device

Recalled March 5, 2025 · FDA recall Z-1238-2025 · Pentax of America

Hazard
During endoscopic procedures using a combination of the video processor EPK-i8020c and the i20c series video endoscope, the observed image can become reddish or dark (1.8%). Users have observed smoke-like steam and noted that the light guide at the tip is hot during use / after the endoscope is removed from the patient, and may cause thermal injury to the patient's mucous membranes. · FDA Class II
Units
94 units
Sold at
Nationwide

Description

During endoscopic procedures using a combination of the video processor EPK-i8020c and the i20c series video endoscope, the observed image can become reddish or dark (1.8%). Users have observed smoke-like steam and noted that the light guide at the tip is hot during use / after the endoscope is removed from the patient, and may cause thermal injury to the patient's mucous membranes.

Product
Pentax Medical Video Processor- Intended to be used with PENTAX Medical endoscopes, monitors and other peripheral devices for endoscopic diagnosis, treatment, and video observation. Model: EPK-i8020c
Manufacturer(s)
Pentax of America Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov