RecallWatchUSA

Home / Recalls / Medical Devices

Percussionaire Corporation recalls Phasitron 5, IPV In-Line Valve, (Intrapulmonary Percussive Ventilation) REF: P5-TEE-20 (pack of 20)/ P5-TEE (individual)

Recalled October 22, 2025 · FDA recall Z-0031-2026 · Percussionaire

Hazard
IFU update; in-line valve, with pressure relief seal may leak when used with ventilator; don't use in neonates/infants, with target tidal volumes less-than50mL or less-than10kg- at high risk of cardiopulmonary/neurological compromise due to unrecognized hypoventilation, updates for continued use in others before design update; leak may cause respiratory acidosis, hypoxemia, or respiratory failure. · FDA Class I
Units
59160
Sold at
Worldwide distribution. US nationwide, Russia, Japan, Switzerland, Canada, Qatar, Chile, South Africa, United Kingdom, Israel, Turkey, and UAE

Description

IFU update; in-line valve, with pressure relief seal may leak when used with ventilator; don't use in neonates/infants, with target tidal volumes less-than50mL or less-than10kg- at high risk of cardiopulmonary/neurological compromise due to unrecognized hypoventilation, updates for continued use in others before design update; leak may cause respiratory acidosis, hypoxemia, or respiratory failure.

Product
Phasitron 5, IPV In-Line Valve, (Intrapulmonary Percussive Ventilation) REF: P5-TEE-20 (pack of 20)/ P5-TEE (individual) with In-Line Valve Sentec Instructions for Use
Manufacturer(s)
Percussionaire Corporation

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Percussionaire Corporation recalls Phasitron 5, IPV In-Line Valve, (Intrapulmonary Percussive Ventilation) REF: P5-TEE-20 (pack of 20)/ P5-TEE (individual) | RecallWatchUSA