RecallWatchUSA

Home / Recalls / Medical Devices

PerkinElmer Health Sciences Canada, Inc. recalls PerkinElmer QSight 210 MD Mass Spectrometer Simplicity 3Q MD v1.0 with Hot Fix 2 Software intended to identify compounds

Recalled December 5, 2018 · FDA recall Z-0513-2019 · PerkinElmer Health Sciences Canada

Hazard
Under certain conditions when using PerkinElmer QSight 210 MD mass spectrometer with Simplicity 3Q MD and Hotfix 2 software, it is possible for the original data from the last acquired sample to be over-written when the automatic flush is enabled. · FDA Class II
Units
18 units
Sold at
FL, MN, PA India, Italy, Korea, Spain

Description

Under certain conditions when using PerkinElmer QSight 210 MD mass spectrometer with Simplicity 3Q MD and Hotfix 2 software, it is possible for the original data from the last acquired sample to be over-written when the automatic flush is enabled.

Product
PerkinElmer QSight 210 MD Mass Spectrometer Simplicity 3Q MD v1.0 with Hot Fix 2 Software intended to identify compounds in human samples by ionizing the compound and separating the resulting ions. Part Number: BC004872
Manufacturer(s)
PerkinElmer Health Sciences Canada, Inc.
Made in
Canada

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov