Home / Recalls / Medical Devices
Perkinelmer recalls Perkin Elmer WIZARD2 10-detector, 1000 samples. Product Code: 2470-0200, 3470-0200. The Wizard gamma counter is intend
Recalled January 20, 2016 · FDA recall Z-0637-2016 · Perkinelmer
Hazard
The content of the second #023 barcode ID label in the Barcode ID Label Binder is incorrect. The second barcode ID label #023 symbol incorrectly identifies as #024 when scanned. If measurement protocols have been programmed for both ID #023 and #024, the protocol associated with barcode ID #024 is executed. If the error is undetected, the Gamma Counter may produce erroneous results. · FDA Class II
Units
22 units
Sold at
Worldwide Distribution - US Nationwide and the countries of ARGENTINA, AUSTRALIA, AUSTRIA, BELGIUM, BOLIVIA, BRAZIL, CANADA, CHINA, DENMARK, FINLAND, FRANCE, GERMANY, GREAT BRITAIN, GREECE, HONG KONG, INDIA, IRELAND, ISRAEL, ITALY, JAPAN, KOREA, KUWAIT, MYANMAR, NETHERLANDS, NEW ZEALAND, NORWAY, POLAND, PORTUGAL, SAUDI ARABIA, SINGAPORE, SLOVAKIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, and THAILAND.
Description
The content of the second #023 barcode ID label in the Barcode ID Label Binder is incorrect. The second barcode ID label #023 symbol incorrectly identifies as #024 when scanned. If measurement protocols have been programmed for both ID #023 and #024, the protocol associated with barcode ID #024 is executed. If the error is undetected, the Gamma Counter may produce erroneous results.
Product
Perkin Elmer WIZARD2 10-detector, 1000 samples. Product Code: 2470-0200, 3470-0200. The Wizard gamma counter is intended to detect and count gamma radiation emitted by clinical samples
Manufacturer(s)
Perkinelmer
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →