Home / Recalls / Medical Devices
PFM MEDICAL INC. recalls Multi-Snare Set: 5 mm x 125 cm, REF: 147305; 10 mm x 125 cm, REF: 147310
Recalled July 1, 2026 · FDA recall Z-2526-2026 · PFM MEDICAL
Hazard
A snare catheter including luer lock, indicated for removal or manipulation of a foreign body in the vascular system, failed to meet the biocompatibility requirements for in-vitro-cytotoxicity, which if in contact with a patient's vasculature could result in localized inflammation and tissue irritation. · FDA Class II
Units
32
Sold at
US Nationwide distribution in the states of NY, MD.
Description
A snare catheter including luer lock, indicated for removal or manipulation of a foreign body in the vascular system, failed to meet the biocompatibility requirements for in-vitro-cytotoxicity, which if in contact with a patient's vasculature could result in localized inflammation and tissue irritation.
Product
Multi-Snare Set: 5 mm x 125 cm, REF: 147305; 10 mm x 125 cm, REF: 147310
Manufacturer(s)
PFM MEDICAL INC.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →