RecallWatchUSA

Home / Recalls / Medical Devices

Phadia Ab recalls EliA SmDP Well, Article Number 14-5624-01 Product Usage: EliA SmDP is intended for the in vitro semi-quantitative me

Recalled February 3, 2016 · FDA recall Z-0697-2016 · Phadia Ab

Hazard
All equivocal and positive results (greater or equal 7 EliA U/ml) on EliA SmDP well lot 0018 may be incorrect and must be considered invalid. Patient samples in complaint investigations caused unspecific signals up to 22 U/ml on EliA SmDP well lot 0018. The nonspecific signals are not caused by anti-Sm antibodies, nor by streptavidin antibodies. · FDA Class II
Units
76
Sold at
US Nationwide in the states of NC, GA, NJ, NY, PA, TX, IL, MI, MT, NM, and OR.

Description

All equivocal and positive results (greater or equal 7 EliA U/ml) on EliA SmDP well lot 0018 may be incorrect and must be considered invalid. Patient samples in complaint investigations caused unspecific signals up to 22 U/ml on EliA SmDP well lot 0018. The nonspecific signals are not caused by anti-Sm antibodies, nor by streptavidin antibodies.

Product
EliA SmDP Well, Article Number 14-5624-01 Product Usage: EliA SmDP is intended for the in vitro semi-quantitative measurement of IgG antibodies directed to Sm in human serum and plasma (EDTA, citrate) as an aid in the clinical diagnosis of systemic lupus erythematosus (SLE) in conjunction with other laboratory and clinical findings. EliA SmDP uses the EliA IgG method on the instruments Phadia 100 and Phadia 250.
Manufacturer(s)
Phadia Ab
Made in
Sweden

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Phadia Ab recalls EliA SmDP Well, Article Number 14-5624-01 Product Usage: EliA SmDP is intended for the in vitro semi-quantitative me | RecallWatchUSA