Philips And Neusoft Medical Systems Co., Ltd. recalls NeuViz 16 Multi-Sliced CT Scanner System, part number 989605858501. NeuViz Dual series CT Scanner System, part number 9

Recalled December 12, 2012 · FDA recall Z-0466-2013 · Philips And Neusoft Medical Systems

Hazard
The four screws that secure the patient table top to the carrier pulled out and caused the table top to detach. This issue may affect NeuViz Dual and NeuViz 16 scanners. · FDA Class II
Units
20 units of NeuViz 16 CT Scanner Systems and 25 units of NeuViz Dual series
Sold at
Nationwide Distribution including North Carolina, Ohio, Texas, and Puerto Rico.

Description

The four screws that secure the patient table top to the carrier pulled out and caused the table top to detach. This issue may affect NeuViz Dual and NeuViz 16 scanners.

Product
NeuViz 16 Multi-Sliced CT Scanner System, part number 989605858501. NeuViz Dual series CT Scanner System, part number 989605651321 The NeuViz 16 CT Scanner System can be used as a whole body computed tomography X-ray system featuring a continuously rotating X-ray tube and detector array with multi-slice capability up to 16 slices simultaneously.
Manufacturer(s)
Philips And Neusoft Medical Systems Co., Ltd.
Made in
China

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov