RecallWatchUSA

Home / Recalls / Medical Devices

Philips Healthcare Inc. recalls Juno DRF , X-ray system, diagnostic, Fluoroscopic, general-purpose Model: 70920 Product Usage: Universal R/F, gener

Recalled February 6, 2013 · FDA recall Z-0702-2013 · Philips Healthcare

Hazard
Values of fluoroscopy time, Air Kerma Rate and Cumulated Air Kerma may not be displayed "at the fluoroscopist's working position" as per 21 CFR 1003.21 · FDA Class II
Units
16 units
Sold at
US Nationwide Distribution including the states of AL, CA, CO, CT, IL, MD, MI, MO, NY, OH, PA

Description

Values of fluoroscopy time, Air Kerma Rate and Cumulated Air Kerma may not be displayed "at the fluoroscopist's working position" as per 21 CFR 1003.21

Product
Juno DRF , X-ray system, diagnostic, Fluoroscopic, general-purpose Model: 70920 Product Usage: Universal R/F, general use
Manufacturer(s)
Philips Healthcare Inc.