Recalled September 26, 2012 · FDA recall Z-2384-2012 · Philips Healthcare
Hazard
Creepage Distance between the two primary phases in the GEO Transformer is 4.5 mm.does not meet (IEC) International standard of 5.5 mm. · FDA Class II
Units
4 systems
Sold at
Worldwide Distribution - USA including OH and WA and the country of Canada
Description
Creepage Distance between the two primary phases in the GEO Transformer is 4.5 mm.does not meet (IEC) International standard of 5.5 mm.
Product
Philips DigitalDiagnost Systems with M Cabinet Model: 712025 Stationary radiographic system, digital-Universal RAD Applications
Manufacturer(s)
Philips Healthcare Inc.