RecallWatchUSA

Home / Recalls / Medical Devices

Philips Healthcare Inc. recalls Philips Easy Diagnost Classic-Stationary x-ray system Catalog Numbers: 70642, 70643, 706030, 706031, 706032 Stationa

Recalled August 22, 2012 · FDA recall Z-2208-2012 · Philips Healthcare

Hazard
Inclination switch in the stand may have been mounted incorrectly, as a result, the switch could not stop the tilting movement of the table base due tol hardware or software errors. · FDA Class II
Units
12 units
Sold at
Worldwide Distribution - USA including IN, MA, MI, MO, NC, NY PA, SC, and WI and to the countries of Canada, HongKong, New Zealand.

Description

Inclination switch in the stand may have been mounted incorrectly, as a result, the switch could not stop the tilting movement of the table base due tol hardware or software errors.

Product
Philips Easy Diagnost Classic-Stationary x-ray system Catalog Numbers: 70642, 70643, 706030, 706031, 706032 Stationary x-ray system, Universal RF system for general use.
Manufacturer(s)
Philips Healthcare Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Philips Healthcare Inc. recalls Philips Easy Diagnost Classic-Stationary x-ray system Catalog Numbers: 70642, 70643, 706030, 706031, 706032 Stationa | RecallWatchUSA