Philips Healthcare Inc. recalls Philips HeartStart XL M4735A Defibrillator/Monitors, used with the M3516A batteries The Heart Start XL Defibrillator
Recalled October 10, 2012 · FDA recall Z-0011-2013 · Philips Healthcare
Hazard
HeartStart XL Defibrillator/Monitor (M4735A) used on a patient when AC power is not in use, with a battery that may fail, the device may shut down unexpectedly without providing the user with a low battery warning message or audible tone. · FDA Class II
Units
2500 units
Sold at
Worldwide distribution: USA (Nationwide) and countries of: AUSTRALIA, BANGLADESH, BELGIUM, BRAZIL, CANADA, CHINA, COLOMBIA, CYPRUS, CZECH REPUBLIC, FINLAND, FRANCE, GERMANY HONG KONG, HUNGARY, INDIA, INDONESIA, IRELAND, ISRAEL, ITALY JAPAN, JORDAN, KAZAKHSTAN, KENYA, KOREA, REPUBLIC OF KUWAIT LEBANON, MACAU, MALAYSIA, MEXICO, NAMIBIA, NETHERLANDS, NORWAY, OMAN, PERU, PHILIPPINES, POLAND, PORTUGAL, ROMANIA RUSSIAN FEDERATION, SAUDI ARABIA, SINGAPORE, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, PROVINCE OF CHINA, THAILAND, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM, and VIET NAM.
Description
HeartStart XL Defibrillator/Monitor (M4735A) used on a patient when AC power is not in use, with a battery that may fail, the device may shut down unexpectedly without providing the user with a low battery warning message or audible tone.
Product
Philips HeartStart XL M4735A Defibrillator/Monitors, used with the M3516A batteries The Heart Start XL Defibrillator/Monitor is a fully featured, defibrillator intended for use by qualified medical personnel, trained in either Advanced Cardiac Life Support or in the operation of the device, in a hospital environment.
Manufacturer(s)
Philips Healthcare Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →