PHILIPS HOME HEALTHCARE SOLUTION recalls ProxiDiagnost N90 Image-intensified fluoroscopic x-ray system
Recalled February 3, 2021 · FDA recall Z-0727-2021 · PHILIPS HOME HEALTHCARE SOLUTION
Hazard
After the user selects the Lock-in command, the kV and mA values are not locked immediately and can adjust if the collimator shutter position is changed directly after the Lock-in command. · FDA Class II
Units
91
Sold at
Nationwide
Description
After the user selects the Lock-in command, the kV and mA values are not locked immediately and can adjust if the collimator shutter position is changed directly after the Lock-in command.
Product
ProxiDiagnost N90 Image-intensified fluoroscopic x-ray system
Manufacturer(s)
PHILIPS HOME HEALTHCARE SOLUTION
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →