Philips Medical Systems (Cleveland) Inc recalls Brilliance BigBore Oncology CT - 728243 Computed Tomography X-ray systems intended to produce cross-sectional Images of
Recalled March 29, 2017 · FDA recall Z-1492-2017 · Philips Medical Systems (Cleveland)
Hazard
When the user performs a retrospective respiratory gated 4D CT scan with the Bellows pulmonary gating device, the CT images might be reconstructed at a single phase, while the annotations on the images incorrectly indicate that the reconstruction is from the requested specific phases (0% to 90%). · FDA Class II
Units
11
Sold at
Worldwide Distribution - US to AR, CA, CO, CT, FL, GA, IL, IN, KY, LA, MA, MD, MI, MO, NC, NE, NJ, NM, NY, OH, TX & WI; Internationally to the following countries: Australia, Austria, Canada, China, Czech Republic, Germany, India, Israel, Jordan, Korea, Mexico, New Cldonia, Romania, Russian Federation, Turkey, & United Kingdom.
Description
When the user performs a retrospective respiratory gated 4D CT scan with the Bellows pulmonary gating device, the CT images might be reconstructed at a single phase, while the annotations on the images incorrectly indicate that the reconstruction is from the requested specific phases (0% to 90%).
Product
Brilliance BigBore Oncology CT - 728243 Computed Tomography X-ray systems intended to produce cross-sectional Images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. These devices may include signal analysts and display equipment, patient and equipment supports, components, and accessories. This information may be used by a trained healthcare professional as a diagnostic tool for the visualization and analysis of anatomical and pathological structures.
Manufacturer(s)
Philips Medical Systems (Cleveland) Inc