Philips Medical Systems (Cleveland) Inc recalls Brilliance CT 16, Brilliance CT 64, Brilliance CT Big Bore Oncology, Brilliance CT Big Bore Radiology, Brilliance iCT, I

Recalled April 23, 2014 · FDA recall Z-1461-2014 · Philips Medical Systems (Cleveland)

Hazard
Block assembly-lock stop subframe of service latch was not manufactured to engineering specifications from the supplier, causing premature fracture of the threaded rod on the patient support service latch. As a result, the table top can become free floating due to the disengaged service latch in the table tops subframe. · FDA Class II
Units
135
Sold at
Nationwide Distribution - USA including the states of CA, FL, IL, IN, KY, MD,SC, and TX.

Description

Block assembly-lock stop subframe of service latch was not manufactured to engineering specifications from the supplier, causing premature fracture of the threaded rod on the patient support service latch. As a result, the table top can become free floating due to the disengaged service latch in the table tops subframe.

Product
Brilliance CT 16, Brilliance CT 64, Brilliance CT Big Bore Oncology, Brilliance CT Big Bore Radiology, Brilliance iCT, Ingenuity CT, Ingenuity Core, Ingenuity Core 128 and Ingenuity Flex scanners are Computed Tomography X-ray systems.
Manufacturer(s)
Philips Medical Systems (Cleveland) Inc