Home / Recalls / Medical Devices
Philips Medical Systems (Cleveland) Inc recalls Brilliance CT 16 Power, Model number 728240 Product Usage: Computed Tomography X-ray systems intended to produce cro
Recalled August 31, 2016 · FDA recall Z-2656-2016 · Philips Medical Systems (Cleveland)
Hazard
After upgrading to 3.6.7 software version via FCO72800643, during reconstruction of gated helical scans, the planned anatomy can be cut off on the CT images. This issue can occur on gated CT helical reconstruction that is not planned at iso-center (0,0). If the region of interest is not visualized in the images, a CT rescan may be performed. · FDA Class II
Units
50
Sold at
US Nationwide Distribution in the states of: AL,AR,AZ,CA,CO,CT,DC,DE,FL,GA,HI,IA,ID,IL,IN,KS,KY,LA,MA, MD,ME,MI,MN,MO,MS,NC,ND,NE,NJ,NM, NV,NY,OH,OK,OR,PA,PR,SC,TN,TX,UT,VA,VT,WA,WI,WV,WY
Description
After upgrading to 3.6.7 software version via FCO72800643, during reconstruction of gated helical scans, the planned anatomy can be cut off on the CT images. This issue can occur on gated CT helical reconstruction that is not planned at iso-center (0,0). If the region of interest is not visualized in the images, a CT rescan may be performed.
Product
Brilliance CT 16 Power, Model number 728240 Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes.
Manufacturer(s)
Philips Medical Systems (Cleveland) Inc
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →