RecallWatchUSA

Home / Recalls / Medical Devices

Philips Medical Systems (Cleveland) Inc recalls Brilliance iCT Computed Tomography X-ray system

Recalled June 10, 2015 · FDA recall Z-1695-2015 · Philips Medical Systems (Cleveland)

Hazard
A calibrated torque wrench malfunctioned at the supplier leading to possible incorrect torque of x-ray tube mounting fasteners. · FDA Class II
Units
2
Sold at
Internationally to the countries of China and Norway.

Description

A calibrated torque wrench malfunctioned at the supplier leading to possible incorrect torque of x-ray tube mounting fasteners.

Product
Brilliance iCT Computed Tomography X-ray system
Manufacturer(s)
Philips Medical Systems (Cleveland) Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov