Philips Medical Systems (Cleveland) Inc recalls FV-RF (Clinix RF Plus) Product Usage: A Stationary X-ray System, a device intended for radiographic/fluoroscopic exa
Recalled January 18, 2017 · FDA recall Z-1005-2017 · Philips Medical Systems (Cleveland)
Hazard
There's a potential for liquid penetration into the back side of the front control panel of the X-ray diagnostic table, which can cause a short circuit of the terminals of the tilt operation and cause an unintentional tilt of the table. · FDA Class II
Units
41
Sold at
US Nationwide distribution in the following states: AL, CA, FL, GA, HI, IL, IN, KS, LA, MA, MI, MS, NJ, NY, OH, PA, SC, VA, WI, WV, & WY.
Description
There's a potential for liquid penetration into the back side of the front control panel of the X-ray diagnostic table, which can cause a short circuit of the terminals of the tilt operation and cause an unintentional tilt of the table.
Product
FV-RF (Clinix RF Plus) Product Usage: A Stationary X-ray System, a device intended for radiographic/fluoroscopic examinations of various anatomical regions. This device may include signal analysis and display equipment, patient and equipment supports, components and accessories.