Philips Medical Systems (Cleveland) Inc recalls IntelliSpace Portal DX/HX/EX AutoSPECT Pro Software Application, Philips Medical Systems, Cleveland, OH. Provides sof
Recalled April 27, 2016 · FDA recall Z-1506-2016 · Philips Medical Systems (Cleveland)
Hazard
The AutoSPECT Pro application was only designed to reconstruct cardiac SPECT data obtained with detectors positioned at 90¿ or 180¿ relative to one another. However, certain gamma cameras allow for other relative detector angles. Data acquired at these other angles will not be correctly reconstructed by AutoSPECT Pro and the results will likely be erroneous. · FDA Class II
Units
33 Units
Sold at
Worldwide Distribution -- US, including the states of CO, DE, FL, GA, IL, IN, KS, MO, NC, NV, NY, OH, OR, PA, TN & TX; and, the countries of Australia, Austria, Belgium, Canada, China, Denmark, Egypt, France, Germany, Greece, Ireland, Israel, Italy, Lebanon, Malaysia, Saudi Arabia, Slovenia, South Africa, Switzerland, Turkey, & United Kingdom.
Description
The AutoSPECT Pro application was only designed to reconstruct cardiac SPECT data obtained with detectors positioned at 90¿ or 180¿ relative to one another. However, certain gamma cameras allow for other relative detector angles. Data acquired at these other angles will not be correctly reconstructed by AutoSPECT Pro and the results will likely be erroneous.
Product
IntelliSpace Portal DX/HX/EX AutoSPECT Pro Software Application, Philips Medical Systems, Cleveland, OH. Provides software applications used to process, analyze, and display medical images/data.
Manufacturer(s)
Philips Medical Systems (Cleveland) Inc