RecallWatchUSA

Home / Recalls / Medical Devices

Philips Medical Systems (Cleveland) Inc recalls Philips Healthcare IQon Spectral CT computed tomography x-ray system

Recalled January 17, 2018 · FDA recall Z-0333-2018 · Philips Medical Systems (Cleveland)

Hazard
There is a possibility that screws holding an internal portion of the gantry (Reaction Ring) to the scanner s main bearing may come loose. If this occurs, the part(s) will not be expelled, however, a noise may be heard coming from the gantry. · FDA Class II
Units
50
Sold at
107 domestic accounts 15 US Govt accounts 682 Foreign Accounts 3 Mexico accounts 12 Canada accounts

Description

There is a possibility that screws holding an internal portion of the gantry (Reaction Ring) to the scanner s main bearing may come loose. If this occurs, the part(s) will not be expelled, however, a noise may be heard coming from the gantry.

Product
Philips Healthcare IQon Spectral CT computed tomography x-ray system
Manufacturer(s)
Philips Medical Systems (Cleveland) Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Philips Medical Systems (Cleveland) Inc recalls Philips Healthcare IQon Spectral CT computed tomography x-ray system | RecallWatchUSA