Home / Recalls / Medical Devices
Philips Medical Systems (Cleveland) Inc recalls Philips Medical Systems, GEMINI TF Big Bore, Model #882476, Positron Emission Tomography (PET) and X-Ray Computed Tomogr
Recalled March 6, 2013 · FDA recall Z-0883-2013 · Philips Medical Systems (Cleveland)
Hazard
Software anomaly. Philips determined that the TumorLOC software program for the referenced PET/CT X-Ray scanning systems has the capability to generate Intensity Projection data sets from respiratory gated data. A circumstance was identified by the firm whereby these generated datasets are 'flipped' and incorrectly labeled left-to-right when the are saved to disk. If the flipped Intensity Project · FDA Class II
Units
37
Sold at
Worldwide Distribution - USA including AK, AK, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, OA MA, MD MI, MN, MO, MS, NC, D, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV; Internationally to Canada and other countries..
Description
Software anomaly. Philips determined that the TumorLOC software program for the referenced PET/CT X-Ray scanning systems has the capability to generate Intensity Projection data sets from respiratory gated data. A circumstance was identified by the firm whereby these generated datasets are 'flipped' and incorrectly labeled left-to-right when the are saved to disk. If the flipped Intensity Project
Product
Philips Medical Systems, GEMINI TF Big Bore, Model #882476, Positron Emission Tomography (PET) and X-Ray Computed Tomography (CT) scanning system, a diagnostic imaging systems for fixed or mobile installations that combines Positron Emission Tomography (PET) and X-ray Computed Tomography (CT) systems.
Manufacturer(s)
Philips Medical Systems (Cleveland) Inc
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →