Philips Medical Systems (Cleveland) Inc recalls Philips MX 16-Slice CT Scanner System, a whole-body computed tomography X-ray system.

Recalled March 6, 2013 · FDA recall Z-0899-2013 · Philips Medical Systems (Cleveland)

Hazard
The four screws that secure the patient table top to the carrier became loose, which caused the table top to detach. · FDA Class II
Units
412 Units
Sold at
Worldwide Distribution - USA including AL, AR, CA, FL, GA, IL, IN, KY, MI, MS, NC, NJ, NV, NY, OH, PA, PR and VA.

Description

The four screws that secure the patient table top to the carrier became loose, which caused the table top to detach.

Product
Philips MX 16-Slice CT Scanner System, a whole-body computed tomography X-ray system.
Manufacturer(s)
Philips Medical Systems (Cleveland) Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov