RecallWatchUSA

Home / Recalls / Medical Devices

Philips Medical Systems Gmbh, DMC recalls CombiDiagnost PCF, Software v. 1.0.5, Product Code 709030 - Product Usage: Xray system, diagnostic, fluoroscopic, genera

Recalled February 26, 2020 · FDA recall Z-1329-2020 · Philips Medical Systems Gmbh, DMC

Hazard
When using the Table Up/Down button, the system may experience Error 80, which locks the geometry in that specific state, requiring the intervention of a service Engineer. Additionally, the thermos switch, which handles power down of the unit in case of transformer overheating, was installed incorrectly at production. · FDA Class II
Units
131
Sold at
Worldwide distribution  US Nationwide distribution in the states of AZ, CA, CO, FL, GA, IL, IN, MI, MN, MO, MT, NC, NJ, NY, OH, TX, UT, VA and countries of Canada, Australia, Austria, Bangladesh, Belgium, Bulgaria, Chile, China, Colombia, C¿te D'Ivoire, Cyprus, Czech Republic, Dominican Republic, Egypt, Ethiopia, France, Germany, Ghana, Hungary, Iran, Italy, Jordan, Kenya, Kuwait, Latvia, Lebanon, Libya, Luxembourg, Malaysia, Martinique, Mexico, Nepal, Netherlands, New Zealand, Nigeria, Oman, Philippines, Poland, Portugal, Romania, Russian Federation, Rwanda, Saudi Arabia, Serbia, South Africa, Spain, Switzerland, Tanzania, Thailand, Turkey, United Arab Emirates, United Kingdom, Viet Nam.

Description

When using the Table Up/Down button, the system may experience Error 80, which locks the geometry in that specific state, requiring the intervention of a service Engineer. Additionally, the thermos switch, which handles power down of the unit in case of transformer overheating, was installed incorrectly at production.

Product
CombiDiagnost PCF, Software v. 1.0.5, Product Code 709030 - Product Usage: Xray system, diagnostic, fluoroscopic, general purpose.
Manufacturer(s)
Philips Medical Systems Gmbh, DMC
Made in
Germany

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Philips Medical Systems Gmbh, DMC recalls CombiDiagnost PCF, Software v. 1.0.5, Product Code 709030 - Product Usage: Xray system, diagnostic, fluoroscopic, genera | RecallWatchUSA