PHILIPS MEDICAL SYSTEMS recalls Ingenuity CT CT5000 Ingenuity Plus CT5000 Ingenuity Pro CT5000 Ingenuity Premium Ingenuity CT Upgrades Ingenuity CT

Recalled August 5, 2026 · FDA recall Z-2652-2026 · PHILIPS MEDICAL SYSTEMS

Hazard
The clinical CT scan may not automatically trigger after the Bolus Tracking scan and may need to be triggered manually. Additional Bolus Tracking shots may occur. This may result in the clinical acquisition being inadequate for diagnosis, causing a potential for patient rescan and receiving additional radiation. · FDA Class II
Units
182
Sold at
US Nationwide Distribution including Puerto Rico

Description

The clinical CT scan may not automatically trigger after the Bolus Tracking scan and may need to be triggered manually. Additional Bolus Tracking shots may occur. This may result in the clinical acquisition being inadequate for diagnosis, causing a potential for patient rescan and receiving additional radiation.

Product
Ingenuity CT CT5000 Ingenuity Plus CT5000 Ingenuity Pro CT5000 Ingenuity Premium Ingenuity CT Upgrades Ingenuity CT China Ingenuity Core 128 Elite Brazil
Manufacturer(s)
PHILIPS MEDICAL SYSTEMS

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov