RecallWatchUSA

Home / Recalls / Medical Devices

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Allura Xper FD20; Model Number: 722028;

Recalled January 21, 2026 · FDA recall Z-1075-2026 · PHILIPS MEDICAL SYSTEMS NEDERLAND B.V

Hazard
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown. · FDA Class II
Units
2 units
Sold at
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Bangladesh, Belgium, Brazil, Cayman Islands, Chile, China, Colombia, Costa Rica, Cuba, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, France, Germany, Greece, Guatemala, Honduras, India, Italy, Japan, Kenya, Korea, Republic of, Malaysia, Mexico, Nepal, Netherlands, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Qatar, Romania, Russian Federation, Saudi Arabia, Singapore, South Africa, Spain, Sri Lanka, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Venezuela, Yemen.

Description

Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.

Product
Allura Xper FD20; Model Number: 722028;
Manufacturer(s)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Made in
Netherlands

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Allura Xper FD20; Model Number: 722028; | RecallWatchUSA