PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Azurion system; System Code Description (Model Numbers): Azurion 3 M12 (722063), Azurion 3 M15 (722064), Azurion 7 B12 (

Recalled May 27, 2026 · FDA recall Z-2227-2026 · PHILIPS MEDICAL SYSTEMS NEDERLAND B.V

Hazard
The deaeration hose in X-ray tube cooling units type CU3101, manufactured between February 2016 and May 2020, may degrade over time, potentially resulting in oil leakage. Oil leakage may affect the cooling performance of the X-ray tube. When the oil flow drops below a certain threshold, the system automatically switches to low-dose fluoroscopy and displays the message Low load fluoroscopy flavor selected: Tube cooler problem . This condition cannot be resolved by restarting the system. · FDA Class II
Units
1651 units (592 US, 1359 OUS)
Sold at
Domestic: Nationwide; International: Canada, Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Brunei Darussalam, Bulgaria, Chile, China, Colombia, Costa Rica, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Ethiopia, Finland, France, French Polynesia, Georgia, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Republic of, Kosovo, Kuwait, Kyrgyzstan, Lao People's Democratic Republic, Latvia, Lebanon, Libya, Lithuania, Malaysia, Maldives, Malta, Martinique, Mexico, Moldova, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Nigeria, Norway, Oman, Pakistan, Palestine, State of, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Federation, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Taiwan, Thailand, Trinidad and Tobago, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Viet Nam, Yemen;

Description

The deaeration hose in X-ray tube cooling units type CU3101, manufactured between February 2016 and May 2020, may degrade over time, potentially resulting in oil leakage. Oil leakage may affect the cooling performance of the X-ray tube. When the oil flow drops below a certain threshold, the system automatically switches to low-dose fluoroscopy and displays the message Low load fluoroscopy flavor selected: Tube cooler problem . This condition cannot be resolved by restarting the system.

Product
Azurion system; System Code Description (Model Numbers): Azurion 3 M12 (722063), Azurion 3 M15 (722064), Azurion 7 B12 (722067), Azurion 7 B20 (722068), Azurion 7 M12 (722078), Azurion 7 M20 (722079, 722224).
Manufacturer(s)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Made in
Netherlands

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov