PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Philips AneurysmFlow. Model Number: 001015. AneurysmFlow is a software medical device (Interventional Tool) intended

Recalled April 22, 2026 · FDA recall Z-1805-2026 · PHILIPS MEDICAL SYSTEMS NEDERLAND B.V

Hazard
Potential safety issue involving the Mean Aneurysm Flow Amplitude (MAFA) ratio of AneurysmFlow (Interventional Tool). Philips has determined that the MAFA ratio does not provide reliable prognostic information regarding aneurysm occlusion following Flow Diverter Stent treatment, however despite being stated in the IUF, it is being relied on for making clinical decisions. · FDA Class II
Units
1,033 units
Sold at
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Azerbaijan, Bahrain, Belarus, Belgium, Bnis insted Us Instas instes, Brazil, Canada, Chile, Colombia, Croatia, Czech Republic, Egypt, Estonia, Germany, Greece, Indonesia, Iraq, Italy, Japan, Kuwait, Latvia, Macao, Maldives, Mongolia, Morocco, New Zealand, Nicaragua, Poland, Portugal, Romania, Saudi Arabia, Serbia, Slovenia, South Africa, Sri Lanka, Switzerland, TURKEY, Ukraine, Utd.Arab.Emir., Uzbekistan, Vietnam.

Description

Potential safety issue involving the Mean Aneurysm Flow Amplitude (MAFA) ratio of AneurysmFlow (Interventional Tool). Philips has determined that the MAFA ratio does not provide reliable prognostic information regarding aneurysm occlusion following Flow Diverter Stent treatment, however despite being stated in the IUF, it is being relied on for making clinical decisions.

Product
Philips AneurysmFlow. Model Number: 001015. AneurysmFlow is a software medical device (Interventional Tool) intended to be used in combination with a Philips interventional X-ray system and 3DRA data.
Manufacturer(s)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Made in
Netherlands

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov