PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Philips Azurion 3M12, Model Numbers: 722063, 722221 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with

Recalled June 10, 2026 · FDA recall Z-2284-2026 · PHILIPS MEDICAL SYSTEMS NEDERLAND B.V

Hazard
During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run. · FDA Class II
Units
16 units
Sold at
US Nationwide distribution.

Description

During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.

Product
Philips Azurion 3M12, Model Numbers: 722063, 722221 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key
Manufacturer(s)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Made in
Netherlands