Home / Recalls / Medical Devices
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Philips Azurion 3M15, Model Numbers: 722064, 722222 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 wit
Recalled June 10, 2026 · FDA recall Z-2285-2026 · PHILIPS MEDICAL SYSTEMS NEDERLAND B.V
Hazard
During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run. · FDA Class II
Units
27 units
Sold at
US Nationwide distribution.
Description
During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.
Product
Philips Azurion 3M15, Model Numbers: 722064, 722222 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key
Manufacturer(s)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Made in
Netherlands
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →