RecallWatchUSA

Home / Recalls / Medical Devices

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Veradius Unity, Model #718132

Recalled February 23, 2022 · FDA recall Z-0649-2022 · PHILIPS MEDICAL SYSTEMS NEDERLAND B.V

Hazard
Firm discovered that the wireless foot switch can suddenly stop responding due to a firmware issue. If the wireless foot switch fails, there could be a delay or interruption of procedure if the hand switch is not used or if the wired foot switch is not connected to the system. · FDA Class II
Units
12 US; 44 ROW
Sold at
Domestic distribution to AZ CA FL GA HI IL KS KY MA MD ME MI MN MS NC NE NJ NV NY OH PA TN TX VA WA WV. Worldwide foreign distribution.

Description

Firm discovered that the wireless foot switch can suddenly stop responding due to a firmware issue. If the wireless foot switch fails, there could be a delay or interruption of procedure if the hand switch is not used or if the wired foot switch is not connected to the system.

Product
Veradius Unity, Model #718132
Manufacturer(s)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Made in
Netherlands

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Veradius Unity, Model #718132 | RecallWatchUSA