RecallWatchUSA

Home / Recalls / Medical Devices

Philips Medical Systems Nederlands recalls Ingenia Elition S, Ingenia Elition X

Recalled January 16, 2019 · FDA recall Z-0694-2019 · Philips Medical Systems Nederlands

Hazard
During automatic tabletop movement, pressing and holding the Stop Table button may lead to an unexpected restart of the table movement. · FDA Class II
Units
35
Sold at
CA, MN, NJ, NY, OH, TN, TX. Foreign distribution to Canada, Belgium, France, Germany, INdia, Japna, Netherlands, South Africa, Sweden, Switzerland, UAE.

Description

During automatic tabletop movement, pressing and holding the Stop Table button may lead to an unexpected restart of the table movement.

Product
Ingenia Elition S, Ingenia Elition X
Manufacturer(s)
Philips Medical Systems Nederlands
Made in
Netherlands

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov