Philips Medical Systems Technologies Ltd. recalls Vue PACS Diagnostic Viewer with Vue PACS client 12.1.5 to 12.2.8.454 software Model Number: 1149418 Intended use is
Recalled November 27, 2024 · FDA recall Z-0497-2025 · Philips Medical Systems Technologies
Hazard
A software issue affecting Philips Vue PACS may cause incorrect ischemic map and table value calculations when using the export function in the Perfusion Application for a Siemens CT perfusion study, resulting in misdiagnosis due to incorrect ischemic map and table values. · FDA Class II
Units
2439 systems
Sold at
US Nationwide and Worldwide Distribution Foreign: Country Argentina Australia Austria Belgium Brazil Chile China Colombia Denmark Finland France Germany Gibraltar Greece Guernsey Holy See (Vatican City State) Hong Kong Hungary Indonesia Ireland Israel Italy Japan Jersey Jordan Lebanon Malaysia Mexico Netherlands New Zealand Norway Oman Panama Philippines Poland Portugal Romania Saudi Arabia Serbia Slovenia Spain Sweden Switzerland Taiwan Thailand United Kingdom
Description
A software issue affecting Philips Vue PACS may cause incorrect ischemic map and table value calculations when using the export function in the Perfusion Application for a Siemens CT perfusion study, resulting in misdiagnosis due to incorrect ischemic map and table values.
Product
Vue PACS Diagnostic Viewer with Vue PACS client 12.1.5 to 12.2.8.454 software Model Number: 1149418 Intended use is to provide completely scalable local and wide area PACS (Picture Archiving and Communication System) solutions for hospital and related institutions/sites.
Manufacturer(s)
Philips Medical Systems Technologies Ltd.
Made in
Israel
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →