Home / Recalls / Medical Devices
Philips North America recalls (1) Ingenia Ambition X (Product Number 781356); (2) Ingenia Ambition S (Product Number 781359)
Recalled January 17, 2024 · FDA recall Z-0715-2024 · Philips North America
Hazard
The sealed magnet will experience uncontrolled shutdown known as Loss-of-Field (LoF) if the magnet loses cryo-cooling (e.g. caused by a power outage on site.) · FDA Class II
Units
56 US; 169 ROW
Sold at
Domestic distribution to the following states and territories: CA DE FL GA IL KY MA ME MI MO NH NJ NM NY OH PA SC TN TX VA VT WI WY, and PR. International distribution worldwide.
Description
The sealed magnet will experience uncontrolled shutdown known as Loss-of-Field (LoF) if the magnet loses cryo-cooling (e.g. caused by a power outage on site.)
Product
(1) Ingenia Ambition X (Product Number 781356); (2) Ingenia Ambition S (Product Number 781359)
Manufacturer(s)
Philips North America
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →