Home / Recalls / Medical Devices
Philips North America recalls Achieva 1.5T Initial system. Model (REF) Numbers 781178.
Recalled February 7, 2024 · FDA recall Z-0947-2024 · Philips North America
Hazard
The Quadrature Body Coil (QBC) seal adhesive may fail, creating sharp edges that may come in contact with patients. If the QBC seal becomes loose during the scanning process, the risk to the patient may include skin abrasions, bruises, lacerations, hair loss/entanglement, and tissue injury. · FDA Class II
Units
23 US; 51 OUS
Sold at
Worldwide - US Nationwide distribution.
Description
The Quadrature Body Coil (QBC) seal adhesive may fail, creating sharp edges that may come in contact with patients. If the QBC seal becomes loose during the scanning process, the risk to the patient may include skin abrasions, bruises, lacerations, hair loss/entanglement, and tissue injury.
Product
Achieva 1.5T Initial system. Model (REF) Numbers 781178.
Manufacturer(s)
Philips North America
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →