Philips North America recalls BrightView XCT, gamma camera for SPECT; Model Nos.: 6-digit format 882482; 12-digit format 453560462131 45356074

Recalled February 14, 2024 · FDA recall Z-0931-2024 · Philips North America

Hazard
Detector may unexpectedly fall due to a component failure. If the detector is positioned below center of gantry, there is a potential for abrasion, contusion, laceration, and/or fracture to the patient's lower limbs, as well as an interruption to normal system operation. If detector is positioned above center of gantry, there may be an interruption to normal system operation. · FDA Class I
Units
106 US; 241 OUS
Sold at
Domestic distribution nationwide. International distribution worldwide.

Description

Detector may unexpectedly fall due to a component failure. If the detector is positioned below center of gantry, there is a potential for abrasion, contusion, laceration, and/or fracture to the patient's lower limbs, as well as an interruption to normal system operation. If detector is positioned above center of gantry, there may be an interruption to normal system operation.

Product
BrightView XCT, gamma camera for SPECT; Model Nos.: 6-digit format 882482; 12-digit format 453560462131 453560749161
Manufacturer(s)
Philips North America

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov