Home / Recalls / Medical Devices
Philips North America recalls IntelliVue G7m Anesthesia Gas Module, Product Number 866173
Recalled October 9, 2024 · FDA recall Z-0007-2025 · Philips North America
Hazard
The 866173 lntelliVue G7m Anesthesia Gas Modules could have a component issue that may trigger an interruption of the gas measurement. This issue concerns the G7m's sample gas pump which could stop working, producing a technical, INOP (inoperable) alarm. The pump may fail prematurely when there is friction caused by an abrasion. This recall is for an IMPORTANT PRODUCT NOTICE dated 1/18/23. · FDA Class II
Units
8,449 devices
Sold at
US Nationwide. Global Distribution.
Description
The 866173 lntelliVue G7m Anesthesia Gas Modules could have a component issue that may trigger an interruption of the gas measurement. This issue concerns the G7m's sample gas pump which could stop working, producing a technical, INOP (inoperable) alarm. The pump may fail prematurely when there is friction caused by an abrasion. This recall is for an IMPORTANT PRODUCT NOTICE dated 1/18/23.
Product
IntelliVue G7m Anesthesia Gas Module, Product Number 866173
Manufacturer(s)
Philips North America
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →