Philips North America Llc recalls Achieva 1.5T Initial system Model Number (REF): 781178

Recalled May 29, 2024 · FDA recall Z-1838-2024 · Philips North America

Hazard
The g-MDU (global Mains Distribution Unit) L3 terminal connection may become loose creating a hotspot that may cause smoke/fire to alarm in the hospital's technical room, resulting in Smoke and/or fire · FDA Class II
Units
2 units
Sold at
Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Canada, China, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Italy, Japan, Lebanon, Libya, Lithuania, Martinique, Netherlands, Norway, Poland, Romania, Slovakia, Spain, Sweden, Switzerland, Thailand, United Kingdom.

Description

The g-MDU (global Mains Distribution Unit) L3 terminal connection may become loose creating a hotspot that may cause smoke/fire to alarm in the hospital's technical room, resulting in Smoke and/or fire

Product
Achieva 1.5T Initial system Model Number (REF): 781178
Manufacturer(s)
Philips North America Llc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov