Philips North America, LLC recalls Allura Exper FD Series. System codes 722003 722010 722022 722005 722011 722019 722001 722002 722006 722012 72

Recalled March 25, 2020 · FDA recall Z-1486-2020 · Philips North America

Hazard
A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a short period of time. When this occurs, no image acquisition is possible. · FDA Class II
Units
4938
Sold at
Nationwide domestic distribution. Foreign distribution worldwide.

Description

A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a short period of time. When this occurs, no image acquisition is possible.

Product
Allura Exper FD Series. System codes 722003 722010 722022 722005 722011 722019 722001 722002 722006 722012 722008 722013 722020 722025 722015 722023 722038 889006 889016 889009 722005 The device is an X-Ray Generator.
Manufacturer(s)
Philips North America, LLC